The airpointer‘s modular design comprising a base unit, analyzing modules and sensor modules allows for a configuration according to different application requirements.
The airpointer‘s compact design enables it to be installed almost everywhere. Due to its optimized thermal management, the airpointer consumes less power compared to conventional monitoring stations.
The airpointer offers a choice of analysis modules using type approved reference methods for monitoring airborne pollutants (SO2, NO2 / NOX, CO, O3, and PM) classified as relevant by the EU, the WHO, the US-EPA and further responsible organizations all over the world.
With our fully integrated gas chromatograph we can measure BTX (Benzene/Toluene/Ethylbenzene/Xylene) with out any carrier gas and provide daily aed calibration check.
A fast optical system or an approved PM analyzer is used for monitoring PM.
The integrated data management system records monitoring data of the airpointer‘s own analysis modules as well as various external third-party sensors. An internal web server enables data retrieval by using any Internet connection. Data are available worldwide via access authorisation and can be presented in clearly arranged graphics. All parameters can be displayed locally or online.
| Parameter | airpointer PM (HC) |
|---|---|
| Standard modules | 4 of the following modules |
| NO/NO₂/NOX | Weight: 12.0 kg / 26.5 lbs — optional: span module |
| O₃ | Weight: 5.8 kg / 12.5 lbs — optional: span module |
| SO₂ (H₂S) | Weight: 8.5 kg / 18.7 lbs — optional: span module, H₂S module |
| CO | Weight: 9.0 kg / 19.8 lbs — optional: span module |
| PM10/PM2.5 | Type approved PM10 and PM2.5 (Met One BAM1020 or EDM 180C) |
| BTX | See page 17 for technical specifications |
| Dimensions (H/W/D, w/o sample inlets) | 1480 × 920 × 650 mm / 58.28 × 36.22 × 25.59 in |
| Weight | 110 kg / 242.5 lbs |
| Power consumption | 2,000 W |
| Flow without Dust | < 3,000 ccm/min |
| Operating temperature | –20 °C / –4 °F to +50 °C (optional heating down to –40 °C / –40 °F) |
| Construction | Well-isolated double aluminum construction |
| Standard equipment | Removable module drawers, rugged burglar-proof design, internal air conditioning & temperature control, maintenance door, cylinder lock, zero air supply |
| Options | Mounting brackets, wireless communication (LTE/UMTS/3G), sample gas conditioning, external device integration (4–20 mA, RS-232, Modbus via IP), external data system communication, various base frames (roadside, indoor, pickup truck, trailer) |
| Parameter | Specification |
|---|---|
| Measured Compound | Nitrogen Oxides NO / NO₂ / NOX |
| EU Directive / USEPA Procedure | Chemiluminescence (EN14211) |
| Measurement Principle | Chemiluminescence |
| Range | Dynamic, up to 20 ppm |
| Zero noise | 0.2 ppb RMS |
| Lower detection limit | 0.4 ppb |
| Zero drift (24 hours) | < 0.4 ppb |
| Span drift (24 hours) | ±1 % of reading > 100 ppb NO |
| Response time | < 60 seconds |
| Precision | 1 % of reading or 1 ppb (whichever is greater) @ < 500 ppb |
| Linearity | ±1 % of reading > 100 ppb NO |
| Sample flow rate | 1,000 ml/min |
| Parameter | Specification |
|---|---|
| Measured Compound | Ozone O₃ |
| EU Directive / USEPA Procedure | UV photometry (EN14625) |
| Measurement Principle | UV photometry |
| Range | Dynamic, up to 20 ppm |
| Zero noise | 0.25 ppb RMS |
| Lower detection limit | 0.5 ppb |
| Zero drift (24 hours) | < 1 ppb |
| Span drift (24 hours) | ±1 % of reading or 1 ppb (whichever is greater) |
| Response time | < 30 seconds |
| Precision | 1 ppb |
| Linearity | ±1 % of reading > 100 ppb O₃ |
| Sample flow rate | approx. 1,000 ml/min |
| Parameter | Specification |
|---|---|
| Measured Compound | Sulfur Dioxide SO₂ and Hydrogen Sulfide H₂S |
| EU Directive / USEPA Procedure | UV Fluorescence (EN14212) — for SO₂ |
| Measurement Principle | UV fluorescence |
| Range | Dynamic, up to 10 ppm |
| Zero noise | 0.25 ppb RMS |
| Lower detection limit | 0.5 ppb |
| Zero drift (24 hours) | < 1 ppb |
| Span drift (24 hours) | ±1 % of reading > 100 ppb SO₂ |
| Response time | < 90 seconds |
| Precision | 1 % of reading or 1 ppb (whichever is greater) |
| Linearity | ±1 % of maximum > 100 ppb SO₂ |
| Sample flow rate | 500 ml/min |
| Parameter | Specification |
|---|---|
| Measured Compound | Carbon Monoxide CO |
| EU Directive / USEPA Procedure | NDIR gas filter correlation (EN14626) |
| Measurement Principle | NDIR gas filter correlation |
| Range | Dynamic, up to 1,000 ppm |
| Zero noise | 0.02 ppm RMS |
| Lower detection limit | 0.04 ppb |
| Zero drift (24 hours) | < 0.1 ppb |
| Span drift (24 hours) | ±1 % of reading > 10 ppm CO |
| Response time | < 60 seconds |
| Precision | ±0.1 ppm |
| Linearity | ±1 % of reading < 1,000 ppm CO |
| Sample flow rate | approx. 500 ml/min |
| Parameter | Specification |
|---|---|
| Measured Compound | Benzene / Toluene / Ethylbenzene / Xylene BTX |
| Analysis time | 15 min (15–60 min on request) |
| Standard Range | 0–100 ppb (extended ranges on request) |
| Detector | PID 10.6 eV |
| Column | MEMS, packed |
| Oven Temperature | Ramp up to 140 °C |
| Carrier gas | Air, on board generated: < 20 ml/min |
| Lower detection limit | < 0.25 ppb BTEX (each) for 15 min analysis time |
| Span drift (24 hours) | < 3 % |
| Sample flow rate | 100–250 ml/min typical |
| Parameter | Specification |
|---|---|
| Measured Compound | Particulate Matter |
| EU Directive / USEPA Procedure | Particle collecting and gravimetric analysis |
| Measurement Principle | Nephelometry |
| Range | Dynamic, up to 2,500 μg/m³ |
| Lower detection limit | < 1 μg/m³ |
| Zero drift (24 hours) | < 1 μg/m³ |
| Span drift (24 hours) | ±1 % of reading |
| Response time | < 60 seconds |
| Precision | 1 μg/m³ |
| Sample flow rate | 2 l/min |
| Parameter | Specification |
|---|---|
| Measured Compound | Particulate Matter |
| Particle size ranges | TSP, PM10, PM4, PM2.5, PM1 |
| Concentration range | 0–1,000 μg/m³ |
| Sensitivity | 0.3 μm |
| Sample duration | 1 minute |
| Sample flow rate | 2 l/min |
| Parameter | Specification |
|---|---|
| Measured Compound | Particulate Matter |
| EU Directive / USEPA Procedure | NDIR gas filter correlation (EN14626) |
| Measurement Principle | NDIR gas filter correlation |
| PM values | PM10, PM2.5 (simultaneous) |
| Range | 0.1 to 1,500 μg/m³ |
| Size channels | 31 channels as basis |
| Count range | 1 to 3,000,000 particles/liter |
| Measurement intervals | From 6 seconds up to one hour averages |
| Reproducibility | ±3 % over the total measuring range |
| Laser wavelength | 660 nm |
| Sample flow | Measurement volume of 1.2 l/min ±5 % aically regulated |
To provide the best possible experience, we use technologies such as cookies to store and/or access device information. Consent to these technologies allows us to process data such as browsing behavior or unique IDs on this website. Not consenting or withdrawing consent may adversely affect certain features and functions.