HORIBA has over 100,000 CEM systems installed and 30 years of quality and experience. That is the base on which HORIBA’s ENDA-5000 series of stack-gas analysis systems are built. These systems have a small footprint and use cross modulated non-dispersive infrared (NDIR) and magneto-pneumatic detectors that are inherently drift-free. The ENDA-5000 series are superior continuous analysis systems that perform reliably for difficult exhaust gas measurements, when measurement errors cannot be tolerated. This CEM series features a intuitive touch panel that makes every operation available with the touch of a single button. The ENDA-5000 series systems are also designed for ease of maintenance. They are ideal for a variety of uses, including emissions monitoring from steam boilers, refuse incinerators, and electric power generation plants to assure pollution standards are being met.
CH4 and N2O measurement are also available. Please contact us for more information.
Model | ENDA-5000 | ||||||
Component*1 | NOx | SO2 | CO | CO2 | O2*2 | ||
Measurement methods | NDIR | NDIR | NDIR | NDIR | Magneto-pneumatic detection | ||
Range*3 | Standard | 200~5000 ppm | 200~5000 ppm | 200~5000 ppm | 5~25 vol% | 10~25 vol% | |
Option | 100 ppm~ | 50 ppm~ | 100 ppm~ | – | – | ||
Range Ratio | Within a factor of 10 | Within a factor of 10 | Within a factor of 10 | Within a factor of 5 | Within a factor of 2.5 | ||
Repeatability | Within 0.5% of full scale/week (with optional range, or during O2 measurement, ±1.0% of full scale/week) | ||||||
Linearity (indicator error) | ±1.0% of full scale | ||||||
Zero drift | ±1.0% of full scale/week (assuming surrounding temperature is maintained within 5 °C) | ||||||
Span drift | ±2.0% of full scale/week (assuming surrounding temperature is maintained within 5°) | ||||||
Response time (Td+T90)*4 | < 50 s from the analyzer inlet | ||||||
Overall interference from co-existing gases | ±2.0% of full scale (within standard range, with standard gas formation) | ||||||
Display | Touch panel LCD (backlight) (four usable lines) | ||||||
Environment | Temperature | -5 to 40 °C (away from direct sunlight and radiation heat )*5 | |||||
Humidity | 85% or less (no condensation) | ||||||
Vibration | 100 Hz, 0.3 m/s2or less | ||||||
Dust | Standard environment or better | ||||||
Measuring Gas Condition | Temperature | 250 °C or lower | |||||
Dust | 0.1 g/Nm3 or less | ||||||
Standard gas composition*6 | NO: 500 ppm or less; NO2: 15 ppm or less; SO2: 1000 ppm or less; SO3: 50 ppm or less; | ||||||
Sampling method | Dry sampling using an electric cooler | ||||||
Sample gas flow | 2.5 L/min~3.0 L/min | ||||||
Sample inlet tube | Teflon tubing (Ø8/Ø6 mm) | ||||||
Sample gas pressure | ±4.9 kPa (three points selected) |
scrollable | |||||
Pressure control | Pressure control uses a regulator and pump; Reduced pressure sampling; Control pressure: -4.9 kPa | ||||||
Output | DC 4 to 20 mA (absolute output) (DC 0 to 16 mA/DC 0 to 1 V/DC 1 to 5V optional) Max. 12 output systems | ||||||
External output | Analysis alerts, analysis warnings, range display, corrections, conservation, purging (option) | ||||||
Correction method | Dry correction, aic correction (correction cycle: 7 days standard, can be adjusted to between 1 and 99 days), manual correction | ||||||
Calibration gas | Zero gas | With measurement method authorization: N2, When there is no measurement method authorization: N2 or ambient air | |||||
O2 carrier gas | Ambient air | ||||||
Span gas | Gas cylinder for each component measured (when there is no measurement method authorization: O2 or ambient air can be used) | ||||||
Probe collection point filter | Flange: JIS 10K, 40 AFF; Sample probe tube length: 1000 mm; Material: SUS-316 stainless steel; | ||||||
Power supply | AC 100 V ±15 V(85 V~115 V), AC 230V ±15 % | ||||||
Power frequency | 50/60 Hz (switchable) | ||||||
Power consumption | About 800 VA (heated piping 30m: +1100 VA; heater in tray: +300 VA) | ||||||
Exterior dimensions/Mass | 600 (W) x 1770 (H) x 300 (D) mm, Approx. 200kg | ||||||
Materials in contract with sample gas | SUS-316 stainless steel, SUS-304 stainless steel, Teflon, polypropylene, polyethylene, | ||||||
Enclosure | Independent outdoor installation | ||||||
Color/Finish | Semi-gloss Munsell 5Y7/1 on all inner and outer surfaces | ||||||
*1: CH4 and N2O measurement are also available. Please contact us for more information.
*2: No carrier gas cylinder is required.
*3: Up to two ranges are supported for each component.
*4: If response time from sample intake at the stack is required, please feel free to consult us.
*5: Support is available for -15 to 40 °C (cold-climate version) and 5 to 50 °C.
*6: An NH3 scrubber is available as an option for cases where a combined gas includes NH3. SO2 measurement corrected for CH4 interference is available for cases where there is CH4 in the sample gas for SO2 measurement. CO measurement corrected for N2O interference is available for cases where there is N2O in the sample gas for CO measurement.
To provide the best possible experience, we use technologies such as cookies to store and/or access device information. Consent to these technologies allows us to process data such as browsing behavior or unique IDs on this website. Not consenting or withdrawing consent may adversely affect certain features and functions.